ENVIRONMENTAL TEST METHODS
M1 TANK PROGRAM
(FOR THE M1 WEAPON SYSTEM ONLY)
1. SCOPE
1.1 Purpose.
This document establishes uniform environmental test methods for determining
the resistance of land mobility vehicle equipment to the effects of natural and
induced environments peculiar to military operations. It includes
environmental test methods that will provide, as much as possible, consistently
reproducible test results. The test methods described herein are intended to
be implemented by reference in contractual documents and apply to all land
mobility vehicle equipment. When it is known that the equipment will encounter
conditions more severe or less severe than the environmental levels stated
herein, the test may be modified by the applicable document (specification,
drawing, Quality Assurance Requirement (QAR), etc).
1.2 General
requirements.
1.2.1 Test conditions. Unless
otherwise specified herein or in the equipment specification, measurements and
tests shall be made at standard ambient conditions. Standard ambient
conditions are:
Temperature 73
± 18 degrees Fahrenheit (°F)
Relative humidity 50
± 30 percent
Atmospheric pressure 28.5
(+2.0, -3.0) inches of Mercury (in. Hg)
1.2.1.1 Tolerances. Unless
otherwise specified, tolerance of test conditions shall be as follows:
a. Air
temperature at the control sensor shall be ±2.5°F. The equipment sensor
response time (T) shall be 20 seconds or less. Temperature gradient across the
cross-sectional area occupied by the test item shall not exceed 0.5°F per foot
in any direction, but never more than 4°F total (equipment nonoperating).
NOTE: (T)
is the time required for the sensing system to respond to 62.3 percent of a
step change in temperature in the measured environment.
b. Pressure:
When measured by devices such as manometers, ±5 percent or 0.059 inch of
mercury, whichever provides the greatest accuracy. When measured by devices
such as ion gages, ±10 percent to 10-1 Torr.
c. Relative
humidity at the control sensor: ±5 percent.
d. Vibration
amplitude: ±10 percent.
e. Vibration
frequency: ±2 percent, or ±1/2 Hertz (Hz) below 25 Hz.
f. Acceleration:
±10 percent.
1.2.1.2 Test apparatus accuracy.
The accuracy of instruments and test equipment used to control or monitor the
test parameters shall be verified and shall satisfy the requirements of MIL-STD-45662 to the satisfaction of the procuring activity. All
instruments and test equipment used in conducting the tests specified herein
shall: conform to laboratory standards whose calibration is traceable to the
prime standards at the National Institute of Standards and Technology; and,
unless otherwise specified in the equipment specification, have an accuracy of
at least one-third the tolerance for the variable to be measured (in the event
of conflict between this accuracy and a requirement for accuracy in any one of
the test methods of this document, the latter shall govern).
1.2.2 Test temperature stabilization.
1.2.2.1 Operating. Unless otherwise
specified herein or in the equipment specification, temperature stabilization
will have been attained when the temperature of the part of the test item
considered to have the longest thermal lag is changing no more than 3.6°F per
hour.
1.2.2.2 Nonoperating. Unless
otherwise specified, temperature stabilization will have been attained when the
temperature of the part of the test item considered to have the longest thermal
lag reaches a temperature within 3.6°F of the prescribed temperature, except
that any critical component (e.g., battery electrolyte for engine starting
test) will be within 1.8°F. When changing temperatures, the temperature of the
chamber air may be adjusted up to 9°F beyond the desired end point for a period
of time of up to 1 hour to reduce stabilization time, provided that the
stabilization requirements of this paragraph are ultimately attained relative
to the specified end point temperature, and provided the extended chamber
temperatures will not cause damage to the test item.
1.2.3 Performance of test.
1.2.3.1 Pretest performance record.
Prior to proceeding with any of the environmental tests, the test item shall be
operated under standard ambient conditions (see 1.2.1) to obtain data for
determining satisfactory operation of the item as specified in the equipment
specification. A record of specific pretest data shall be made to verify that
the test item performed within the equipment specification requirements.
NOTE: It shall be permissible during consecutive
environmental tests to use post environmental performance test data (see
1.2.3.4) as pretest performance test data as long as the item under test has
not been mishandled and was stored under standard ambient conditions as defined
in 1.2.1.
1.2.3.2 Installation of test items
in test facility. Unless otherwise specified, the test item shall be
installed in the test facility in a manner that will simulate service usage,
making connections and attaching instrumentation as necessary. Plugs, covers,
and inspection plates not used in operation, but used in servicing shall remain
in place. When mechanical or electrical connections are not used, the
connections normally protected in service shall be adequately covered. For
tests where temperature values are controlled, the test chamber shall be at
standard ambient conditions when the test item is installed. The test item may
then be operated to determine that no malfunction or damage was caused due to
faulty installation or handling when operation is required during exposure to
the specified test.
1.2.3.3 Performance check during
test. When operation of the test item is required during the test
exposure, suitable tests shall be performed to determine whether the test
exposure is producing changes in performance when compared with pretest data.
1.2.3.4 Post-test data.
1.2.3.4.1 Preproduction
(qualification) and initial production (first article) testing. At the
completion of each environmental test, the test item shall be inspected for
damage and tested in accordance with the equipment specification. The test
results (data) shall be compared with the pretest data obtained in accordance
with 1.2.3.1.
1.2.3.4.2 Quality conformance
(control) testing. At the completion of all environmental testing, the
test item shall be inspected for damage and performance tested. Performance
testing between environments shall be at the option of the contractor.
1.2.3.5 Failure criteria.
a. The
item shall have failed the test when any of the following occur:
(1) Monitored functional parameters deviate
beyond acceptable limits established in 1.2.3.1.
(2) Catastrophic or structural failure.
(3) Mechanical binding or loose parts, including
screws, clamps, bolts, and nuts, that clearly result in component failure or a
hazard to personnel safety.
(4) Malfunction.
(5) Degradation of performance beyond pretest
record or equipment specification requirements established in 1.2.3.1 (record
to be made after test).
NOTE: Certain types of equipment (e.g., propellants
and electrically driven devices) are often expected to demonstrate lesser
performance at an environmental extreme, particularly low temperature. A
failure would occur only if degradation is more than expected.
(6) Any additional deviations from acceptable
criteria established before the test and recorded according to 1.2.3.1.
(7) Deterioration, corrosion, or change in
tolerance limits of any internal or external parts which could in any manner
prevent the test item from meeting operational service or maintenance requirements.
b. Any
additional or different failure criteria shall be as specified in the equipment
specification.
1.2.3.6 Under/over test. If an
under test or an over test condition occurs, testing shall immediately cease
and direction as to how to proceed shall be obtained from the procuring
activity.
1.2.4 Test facilities and apparatus.
Test facilities, chambers, and apparatus used in conducting the tests contained
in this standard shall be capable of meeting the conditions required.
1.2.4.1
Test chamber. The following shall apply to test chambers, as
applicable.
1.2.4.1.1 Volume. The volume of
the test chamber shall be such that the bulk of the item under test will not
interfere with the generation and maintenance of the test conditions. When testing
multiple sample items simultaneously, the test chamber shall be of sufficient
size so that each test unit is provided uniform environmental conditions and is
not subjected to nontest environments.
1.2.4.1.2 Heat source. The heat
source of the test facility shall be so located that radiant heat from the
source will not fall directly on the test item, except where application of
radiant heat is one of the test conditions.
1.2.4.1.3 Temperature sensors
location. Unless otherwise specified, thermocouples or equivalent
temperature sensors utilized to determine or control the specified chamber
temperature shall be located centrally within the chamber, in the supply
airstream, or in the return airstream whichever provides the specified test
conditions at the bulk under test and shall be baffled or otherwise protected
against radiation effects.
1.2.4.1.4 Internal air circulation.
The conditioned air flow shall be suitably baffled to provide uniform air flow
around the test item. If multiple test items are tested, they shall be so
spaced as to provide free circulation between the test items and the chamber
walls.
1.2.5 Test data. Test data
shall include complete identification of all test equipment and accessories.
The data shall include the actual test sequence used and ambient test
conditions recorded periodically during the test period. Where measureable,
quantitative data is involved, the actual measured value shall be annotated;
checking a box to denote a measurement was within a range, for example, shall
not be acceptable. The test record shall contain a signature and data block
for certification of the test data by the test engineer.
2. APPLICABLE
DOCUMENTS.
2.1 Government
documents.
2.1.1 Specifications, standards, and
handbooks. The following documents of the issue listed in the current
Department of Defense Index of Specifications and Standards (DODISS) and
supplement thereto, if applicable, form a part of this standard to the extent
specified herein.
FEDERAL
P-C 437 Cleaning,
Compound, High Pressure (Steam) Cleaner
MILITARY
MIL-C-9435 Chamber,
Explosion-Proof Testing
MIL-T-43566 Tape, Textile, Cotton or Polyester, General
Purpose, Natural or in Colors
STANDARDS
MIL-STD-45662 Calibration
System Requirements
OTHER
AR 70-38 Research,
Development, Test, and Evaluation of Material for Extreme Climatic Conditions
(Copies of specifications, standards,
handbooks, drawings, and publications required by manufacturers with respect to
specific acquisition functions should be obtained from the contracting activity
or as directed by the contracting officer.)
2.2 Other
publications. The following documents form a part of this standard to the
extent specified herein. Unless otherwise stated, the issue in effect on date
of invitation for bids or request for proposal shall apply.
American Geophysical Union
"The
Relationship of Raindrop Size to Intensity" - Laws and Parsons,
Translations of the American Geophysical Union, Part II, para 452-459, 1943.
(Copies of the
above publication may be obtained from the American Geophysical Union, 2100
Pennsylvania Ave., N.W., Suite 435, Washington, D.C. 20037).
American
Society for Testing Materials (ASTM)
ASTM D910-88
(Copies of the
above publication may be obtained from ASTM, 1916 Race Street, Philadelphia,
PA 19103).
Society of Automotive
Engineers (SAE).
SAE
J726 Air Cleaner Test Code
(Copies of the
above publication may be obtained from SAE, 400 Commonwealth Drive, Warrendale,
PA 15096.
-
3. TEST METHODS.
3.1 Temperature.
3.1.1 High. The high temperature
test is conducted to determine the resistance of equipment to elevated
temperatures that may be encountered during service life either in storage
(without protective packaging) or under service conditions.
3.1.1.1 General effects. High
temperature conditions may cause the permanent set of packings and gaskets,
change the physical and functional characteristics of electrical/electronic
parts, and shorten equipment operating life. In items of complex construction,
binding of parts may also result due to differential expansion of dissimilar
materials. Protective coatings may discolor, materials may crack, closure and
sealing strips may partially melt and adhere to contacting parts. These are a
few of the problems associated with high temperature.
3.1.1.2 Equipment specification data
required. Before proceeding, verify that the equipment specification
specifies the following as a minimum: required pretest data; failure criteria;
highest operating temperature expected for equipment operation; required
measurements; and internal chamber temperature dwell time if other than 48
hours.
3.1.1.3 Procedures. Using a
temperature chamber, proceed as follows:
NOTE: The
rate of temperature change (steps b, d, and f) may be the maximum attainable by
the chamber, but shall not exceed 18°F per minute.
a. Prepare
the test item in accordance with the general requirements of 1.2.
b.
Raise the internal chamber temperature to 160°F, or as specified in the
equipment specification.
c.
Maintain the internal chamber temperature for a period of 48 hours, or as
specified in the equipment specification, while ensuring the relative humidity
is not in excess of 15 percent. If operation of the equipment "during
exposure" is not required, immediately proceed to step f.
d. Adjust the
internal chamber temperature to the highest operating temperature for which
the test item is designed to operate and maintain until temperature stabilization
of the test item is reached.
e. Operate the
test item until the item is stabilized, or as specified in the equipment
specification, and obtain results in accordance with 1.2.3.3.
f. Return the
test item, nonoperating, to standard ambient condition and stabilize.
g. Operate and
inspect test item and obtain results in accordance with 1.2.3.4.
3.1.2 Low. The low temperature
test is conducted to determine the effects of low temperature on equipment
during storage (without protective packaging) and service use.
3.1.2.1 General effects.
Differential contraction of metal parts, loss of resiliency of packing and
gaskets, congealing of lubricants, and changes in the physical and functional
characteristics of electrical/electronic parts are a few of the problems
associated with low temperature.
3.1.2.2 Equipment specification data
required. Before proceeding, verify that the equipment specification
specifies the following as a minimum: required pre-test data; failure
criteria; lowest temperature expected for equipment operation; required
measurement; and internal dwell time.
3.1.2.3 Procedure. Using a
temperature chamber, proceed as follows:
NOTE: The rate
of temperature change (steps b, d, and f) may be the maximum attainable by the
chamber, but shall not exceed 18°F per minute.
a. Prepare
the test item in accordance with the general requirements of 1.2.
b. Lower
the internal chamber temperature to the storage temperature of -60°F, or as
otherwise specified in the equipment specification.
c. Maintain
the temperature for a period of 24 hours after stabilization, or for the period
specified in the equipment specification. If operation of the equipment
"during exposure" is not required, immediately proceed to step f.
d. Adjust
the internal chamber temperature to the lowest temperature under which the test
item is designed to operate as specified in the equipment specification and
maintain until temperature stabilization of the test item is reached.
e. Operate
the test item until the item is stabilized or for the time specified in the
equipment specification and obtain results in accordance with 1.2.3.3.
f. Return
the test item, nonoperating, to standard ambient conditions and stabilize.
g. Operate
and inspect test item and obtain results in accordance with 1.2.3.4.
3.2 Vibration.
The vibration test is performed to determine if equipment is constructed to
withstand expected dynamic vibrational stresses and to ensure that performance
degradations or malfunctions will not be produced by the service vibration
environment. Tests specified herein are established for equipment which is
used in a tank environment.
3.2.1 Equipment specification data
required. Before proceeding, verify that the equipment specification
specifies the following as a minimum: pretest data required; failure criteria;
weight of test item; vibration frequency, amplitude and duration; and
temperature extremes, if applicable.
3.2.2 Procedures. The entire
sequence of tests may be accomplished for any one axis before changing to the
next axis. The transverse motion at the input monitoring point(s) shall be
minimized, and should be limited to 100 percent of the input motion except that
reaction machines shall be balanced to reduce transverse motion +10 percent.
3.2.2.1 Test item operation.
Unless otherwise specified in the equipment specification, the test item shall
be operated during application of vibration so that functional effects caused
by these tests may be evaluated. The procedure provides for a functional and
an endurance vibration test. The test item shall meet the requirements using
the method of test specified in Section 4 of the equipment specification and,
as applicable, the General Requirements of 1.2 herein.
3.2.2.2 Mounting techniques. The
test item shall be attached by its normal mounting means, either directly to
the vibration exciter or transition table, or by means of a rigid fixture
capable of transmitting the vibration conditions specified herein. Precautions
shall be taken in the establishment of mechanical interfaces to minimize the
introduction of extraneous responses in the test setup. The test load shall be
distributed as uniformly as possible on the vibration exciter table in order to
minimize effects of unbalanced loads. The input control sensing device(s)
shall be rigidly attached to the vibration table, or fixture if used, as near
as possible to the attachment point(s) of the test item. Locations to be
selected should include main structure, printed circuit boards, large
components and modules, where practicable. The sensor sizes and weights shall
be limited so that their effect on the dynamic responses being measured is
minimal. For sinusoidal vibration, when necessary for obtaining uniform
results, a tracking filter should be used in the vibration exciter control
feedback loop prior to the servo input.
3.2.2.3 Combined temperature-vibration
test. Tests shall be performed under room ambient conditions unless a high
or low temperature vibration test is specified, in which case the temperature
extremes shall be as specified in the equipment specification.
3.2.2.4 Sinusoidal vibration tests.
The vibration shall be applied along each of three mutually perpendicular axes
of the test item. The vibratory acceleration levels or double amplitudes of
the specified test curve shall be maintained at the test item mounting points.
When necessary for obtaining uniform results, a tracking filter should be used
in the exciter control feedback loop prior to the servo input. When the input
vibration is measured at more than one control point, the control signal shall
be the average of all the accelerometers unless otherwise specified. For
massive test items, fixtures and large force exciters, it is recommended that
the input control level be an average of at least three or more inputs.
3.2.2.5 Cycling. The test item
shall be vibrated along each axis in accordance with the test level and
frequency range of the equipment specification and Figure 1. Unless otherwise
specified in the equipment specification, the frequency of applied vibration
shall be swept over a 5 to 500 to 5 Hertz (Hz) range logarithmically in
accordance with Figure 2 for a period of 180 minutes in each of the three
mutually perpendicular axes. The specified sweep time is that of an ascending
plus a descending sweep and is twice the ascending sweep time shown on Figure 1
for the specified range. When required by the equipment specification, the
test item shall be operating during the test and performance shall be verified
in accordance with 1.2.3.3, as required. At the conclusion of the test, the
test item shall be returned to standard ambient conditions (see 1.2.1), then
examined and tested in accordance with 1.2.3.4.
3.3 Shock. The
shock test is performed to determine if equipment is constructed to withstand
expected dynamic shock stresses and that performance degradations or
malfunctions will not be produced by the service shock environment expected in
handling, transportation, and service use.
3.3.1 Equipment specification data
required. Before proceeding, verify that the equipment specification
specifies the following as a minimum: pretest data required; shock pulse type,
amplitude and duration; temperature extremes, if applicable; and operational
requirements during exposure (i.e., intermittent operation, relay chatter,
etc.).
3.3.2 Apparatus.
3.3.2.1 Shock machine. The shock
machine utilized shall be capable of producing the specified input shock pulse
shown on Figure 3. The shock machine may be of the free fall, resilient
rebound, nonresilient, hydraulic, compressed gas, computer-controlled
electrodynamic exciter or other activating types.
|
|
 |


3.3.2.1.1 Shock machine
calibration. The actual test item, a rejected item, or a rigid dummy mass
shall be used to calibrate the shock machine for conformance with the specified
wave shape. When a rigid dummy mass is used, it shall have the same center of
gravity and the same mass as that intended for the test item and shall be
installed in a manner similar to that of the test item. (When a rigid dummy
mass or rejected item is used for calibration, the waveform during the actual
test may be somewhat different from that observed during calibration.) The
shock machine shall then be calibrated for conformance with the specified
waveform. Two consecutive shock applications to the calibration load shall
produce waveforms which are all within the tolerance envelope given. The
calibrating load shall then be removed and the shock test performed on the
actual test item. Provided all conditions remain the same, other than the
substitution of the test item for the calibrating load, the waveform shall be
considered to meet the specified test requirement. The actual test waveform
shall be recorded for later use should a failure analysis be required.
3.3.2.2 Instrumentation. The
instrumentation used to measure the input shock pulse, in order to meet the
tolerance requirements of the test procedure, shall have the following
characteristics:
3.3.2.2.1 Frequency response.
The frequency response of the complete measuring system, from the accelerometer
through the readout instrument, shall be as shown on Figure 4. Particular care
shall be exercised in the selection of each individual instrument of the shock
measuring instrumentation system in order to assure compatibility with the
prescribed frequency response tolerance.
3.3.2.2.2 Accelerometer,
piezoelectric. When a piezoelectric accelerometer is employed as the shock
sensor, the fundamental resonant frequency of the accelerometer shall be
greater than 14,000 Hz (resonant frequencies of 30,000 Hz or higher are
recommended). For suitable low frequency response, the accelerometer and load
(cathode follower, amplifier, or other load) shall have the following
characteristics:
RC
> 0.2
Where R = load resistance
(ohms)
C = accelerometer
capacitance plus shunt capacitance
of cable
and load (farads)
3.3.2.2.3 Accelerometer, strain gage.
A strain gage accelerometer may be used provided the undamped natural frequency
is equal to or greater than 1,500 Hz with damping approximately 0.64 to 0.70 of
critical.
3.3.2.2.4 Accelerometer calibration.
The accelerometer shall be dynamically calibrated to the specified accuracy.
3.3.2.3.5 Accelerometer mounting.
The monitoring accelerometer shall be rigidly attached to the test item support
fixture at or near the attachment point(s) of the test item.
3.3.3 Procedures.
3.3.3.1 Shock pulse. The shock
pulses shall be as shown on Figure 3. All points of the acceleration waveform
obtained shall lie within the area enclosed by the tolerance limit lines.
3.3.3.2 Mounting of test item.
The test item shall be rigidly attached to the shock machine table following
the general requirements of 1.2.3.2, as applicable. Wherever possible, the
test load shall be distributed uniformly on the test platform in order to
minimize the effects of unbalanced loads.
3.3.3.3 Procedure. This
procedure shall be used for shock testing of equipment assemblies (mechanical,
electrical, hydraulic, electronic, etc.) of medium size, including items which
mount on vibration isolators and equipment racks. Three shocks in each
direction shall be applied along three mutually perpendicular axes of the test
item (total of 18 shocks). If the test item is normally mounted on vibration
isolators, the isolators shall be functional during the test. The shock pulse
shape shall be in accordance with Figure 3 and of duration and amplitude as
specified in the equipment specification. When required by the equipment
specification, the test item shall be operating during the test and performance
verified in accordance with 1.2.3.3, as required. At the conclusion of the
test, the test item shall be operated and inspected and results obtained shall
be verified in accordance with 1.2.3.4.
3.4 Humidity.
The humidity test is conducted to determine the resistance of equipment to the
effects of exposure to warm, highly humid atmosphere such as is encountered in
tropical areas. This is an exaggerated environmental test, accomplished by the
continuous exposure of the equipment to high relative humidity at cycling
elevated temperatures. These conditions impose a vapor pressure on the equipment
under test which constitutes the major force behind the moisture migration and
penetration.
3.4.1 Apparatus. A humidity
chamber is required. The chamber and accessories shall be constructed and
arranged in such a manner as to avoid condensate dripping on the test item.
The chamber shall be trap-vented to the atmosphere to prevent the buildup of
total pressure. Relative humidity shall be determined from the dry bulb-wet
bulb thermometer comparison method or an equivalent method approved by the
procuring activity. When readout charts are used, they shall be capable of
being read with a resolution within 1°F. When the wet bulb control method is
used, the wet bulb and tank shall be cleaned and a new wick installed at least
every 30 days. The air velocity flowing across the wet bulb shall be not less
than 900 feet per minute. Provisions shall be made for controlling the flow of
air throughout the internal chamber test space where the velocity of air shall
not exceed 150 feet per minute. Steam or distilled, demineralized, or
deionized water having a pH

3.4.2 Procedure.
a. Prepare
the test item, as required, in accordance with the general requirements of 1.2
except make initial measurements as specified in step d following.
b. Dry
the test item at 129°F for 24 hours.
c. Condition
the test item at 73 + 18°F and 50 + 10 percent relative humidity
for 24 hours.
d. If
no pre-test data exists, take initial measurements in accordance with 1.2.3.1.
e. Raise
the internal chamber temperature to 86°F and the relative humidity to 95 +
5 percent.
f. Subject
the test item to five continuous 48-hour cycles in accordance with Figure 5.
During the 24th to 28th or 44th to 48th hours of the fifth cycle, operate the
equipment as specified in the equipment specification and verify performance in
accordance with 1.2.3.3, as required.
g. After
completion of step f, condition the test item for 24 hours at 73°F and 50 ± 10
percent relative humidity. During conditioning, remove accumulated moisture by
turning the assembly upside down and shaking, if practicable. In the case of
surfaces such as glass through which one must view, wiping and patting dry is
also acceptable. Disassembly of the unit is not permissible. Any evidence of
internal moisture, visible for example in the case of optical devices, is
unacceptable.
h. Following
conditioning, operate and inspect the item in accordance with the equipment
specification and verify the results in accordance with 1.2.3.4. Evidence of
any internal moisture is cause for rejection.

3.5 Salt fog.
The salt fog test is conducted to determine the resistance of equipment to the
effects of a salt atmosphere. The specified concentration of moisture and salt
is greater than is found in service. The test is applicable to any equipment
exposed to salt fog conditions in service. The test is valuable for
determining the durability of coatings and finishes exposed to a corrosive salt
atmosphere. For other applications, however, this test should be applied only
after full recognition of its deficiencies and limitations.
3.5.1 General effects.
3.5.1.1 Deficiencies.
Testing deficiencies include:
a. The
successful withstanding of this test does not guarantee that the test item will
prove satisfactory under all corrosive conditions.
b. The
salt fog used in this test does not truly duplicate the effects of a marine
atmosphere.
c. It
has not been demonstrated that a direct relationship exists between salt fog
corrosion and corrosion due to other media.
d. This
test is generally unreliable for comparing the corrosion resistance of
different materials or coating conditions, or for predicting their comparative
service life.
3.5.1.2 Limitations. The
salt fog test is acceptable for evaluating the uniformity (i.e., thickness and
degree of porosity) of protective coatings, metallic and nonmetallic, of
different lots of the same product, once some standard level of performance has
been established. When used to check the porosity of metallic coatings, the
test is more dependable when applied to coatings which are cathodic rather than
anodic toward the basic metal. This test can also be used to detect the
presence of free iron contaminating the surface of another metal by inspection
of the corrosion products.
3.5.2 Equipment specification
data required. Before proceeding, verify that the equipment specification
specifies the following as a minimum: pretest data required; failure criteria;
applicable salt solution, if other than 5 percent; salt fog exposure period if
other than 48 hours; drying period if other than 48 hours; inspection and
operation after 24 hours of exposure where buildup of salt deposits are
detrimental to the proper operation of the equipment; and specify whether
operation of an electrical system is required.
3.5.3 Apparatus. The
apparatus used in the salt fog test shall include the following:
a. Exposure
chamber with racks for supporting test items.
b. Salt
solution reservoir with means for maintaining an adequate level of solution.
c. Means
for atomizing salt solution, including suitable nozzles and compressed air
supply.
d. Chamber
heating means and control.
e. Means
for humidifying the air at a temperature above the chamber temperature.
3.5.3.1 Chamber. The chamber
and all accessories shall be made of material that will not affect the corrosiveness
of the fog (e.g., glass, hard rubber, plastic, or kiln dried wood other than
plywood). In addition, all parts which come in contact with test items shall
be of materials that will not cause electrolytic corrosion. The chamber and
accessories shall be constructed and arranged so that there is no direct
impingement of the fog or dripping of the condensate on the test items, that
the fog circulates freely about all test items to the same degree, and that no
liquid which has come in contact with the test items returns to the
salt-solution reservoir. The chamber shall be properly vented to prevent
pressure buildup and allow uniform distribution of salt fog. The discharge end
of the vent shall be protected from strong drafts which can create strong air
currents in the test chamber.
3.5.3.2 Atomizers. The
atomizers used shall be of such design and construction as to produce a finely
divided, wet, dense fog. Atomizing nozzles shall be made of material that is
nonreactive to the salt solution.
3.5.3.3 Air supply. The
compressed air entering the atomizer shall be essentially free from all
impurities, such as oil and dirt. Means shall be provided to humidify and warm
the compressed air as required to meet the operating conditions. The air
pressure shall be suitable to produce a finely divided dense fog with the
atomizer or atomizers used. To insure against clogging the atomizers by salt
deposition, the air should have a relative humidity of at least 85 percent at
the point of release from the nozzle. A satisfactory method is to pass the air
in very fine bubbles through a tower containing heated water which should be
automatically maintained at a constant level. The temperature of the water
should be at least 95°F. The permissible water temperature increases with
increasing volume of air and with decreasing insulation of the chamber and the
chamber's surroundings. However, the temperature should not exceed a value
above which an excess of moisture is introduced into the chamber (for example
109° at an air pressure of 12 psi) or a value which makes it impossible to meet
the requirements for operating temperature.
3.5.3.4 Preparation of salt
solution. The salt used shall be sodium chloride containing, on the dry
basis, not more than 0.1 percent sodium iodide and not more than 0.5 percent of
total impurities. A 5 + 1 percent solution shall be prepared by
dissolving five parts by weight of salt in 95 parts by weight of distilled or
demineralized water. The solution shall be adjusted to and maintained at a
specific gravity between the limits shown on Figure 6 by utilizing the measured
temperature and density of the salt solution. Sodium tetraborate (common
borax) may be added to the salt solution in a ratio not to exceed 0.7 gms (1/4
level teaspoon) sodium tetraborate to 20 gallons of salt solution as a pH
stabilization agent.
a. Adjustment
of pH of the salt solution shall be so maintained that the solution atomized at
95°F and collected by the method specified in 3.5.4.3 will be in the pH range of
6.5 to 7.2.
b. Only
diluted, chemically pure, hydrochloric acid or chemically pure, sodium
hydroxide shall be used to adjust the pH. The addition of sodium tetraborate
as recommended herein will aid in maintaining a stable pH value.
c. The
pH measurement shall be made electrometrically using a glass electrode with a
saturated potassium chloride bridge, or by a colorimetric method, such as
bromothymol blue, provided the results are equivalent to those obtained with
the electrometric method.
d. The
pH shall be measured when preparing each new batch of solution and as specified
in 3.5.4.4.
3.5.3.5 Filter. A filter
fabricated of noncorrosive materials similar to that shown in Figure 7 shall be
provided in the supply line and immersed in the salt solution reservoir in a
manner such as that illustrated in Figure 8.
3.5.4 Procedure.
3.5.4.1 Temperature. The test
shall be conducted with a temperature in the exposure zone maintained at 95°F.
Satisfactory methods for controlling the temperature accurately are by housing
the apparatus in a properly controlled constant temperature room, by thoroughly
insulating the apparatus and preheating the air to the proper temperature prior
to atomization, or by jacketing the apparatus and controlling the temperature
of the water or of the air used in the jacket. The use of immersion heaters
within the chamber for the purpose of maintaining the temperature within the
exposure zone is prohibited.
3.5.4.2 Atomization. Suitable
atomization has been obtained in chambers having a volume of less than 12 cubic
feet under the following conditions:
a. Nozzle
pressure shall be as low as practicable to produce fog at the required rate.
b. Orifices
between 0.02 and 0.03 inch in diameter.
c. Atomization
of approximately 3 quarts of salt solution per 10 cubic feet of chamber volume
per 24 hours.

When using large size chambers having a volume
considerably in excess of 12 cubic feet, the conditions specified may require
modification to meet the requirements for operating conditions.
3.5.4.3 Placement of salt fog
collection receptacles. The salt fog conditions maintained in all parts of
the exposure zone shall be such that a clean fog collecting receptacle placed
at any point in the exposure zone will collect from 0.5 to 3.0 milliliters of
solution per hour for each 80 square centimeters of horizontal collecting area
(10 centimeter diameter) based on an average test of at least 16 hours. A
minimum of two receptacles shall be used, one placed nearest to any nozzle and
one farthest from all nozzles. Receptacles shall be placed so that they are
not shielded by test items and so no drops of solution from test items or other
sources will be collected.
3.5.4.4 Measurement of salt solution.
The solution, collected in a manner specified in 3.7.4.3, shall have the sodium
chloride content and pH specified in 3.5.3.4 when measured at a temperature of
95°F. The salt solution from all collection receptacles used can be combined
to provide that quantity required for the measurements specified.
3.5.4.4.1 Measurement of sodium
chloride content. The solution, maintained at the specified temperature,
can be measured in a graduate of approximately 2.5 centimeters inside
diameter. A small laboratory type hydrometer will be required for measurement
within this volume.
3.5.4.4.2 Time of measurements.
The measurement of both sodium chloride content and pH shall be made at the
following specified times:
a. For
salt fog chambers in continuous use, the measurements shall be made following
each test.
b. For
salt fog chambers that are used infrequently, a 24-hour test run shall be
accomplished followed by the measurements. The test item shall not be exposed
to this test run.
3.5.4.5 Preparation of test item.
The test item shall be given a minimum of handling, particularly on the
significant surfaces, and shall be prepared for test immediately before
exposure. Unless otherwise specified, uncoated metallic or metallic coated
devices shall be thoroughly cleaned of oil, dirt, and grease as necessary until
the surface is free from water break. The cleaning methods shall not include
the use of corrosive solvents nor solvents which deposit either corrosive or
protective films, nor the use of abrasives other than a paste of pure magnesium
oxide. Test items having an organic coating shall not be solvent cleaned.
Those portions of test items which come in contact with the support - and
unless otherwise specified in the case of coated devices or samples, cut edges
and surfaces not required to be coated - shall be protected with a suitable
coating of wax or similar substance impervious to moisture.
|
|
 |
3.5.4.6 Performance of test.
The test item shall be placed in the test chamber in accordance with the
general requirements of 1.2.3.2, and exposed to the salt fog for a period of 48
hours, or as specified in the equipment specification. At the end of the
exposure period, unless otherwise specified, the test item shall be operated
and the results compared with the data obtained in accordance with the general
requirements of 1.2.3.1. The test item shall be inspected for corrosion in
accordance with the general requirements of 1.2.3.4. If necessary to aid in
examination, a gentle wash in running water not warmer than 100°F may be used.
The test item shall then be stored in an ambient atmosphere for 48 hours or as
specified in the equipment specification for drying. At the end of the drying
period, when specified, the test item shall be again operated and the results
compared with the data obtained in accordance with the general requirements of
1.2.3.1. The test item shall then be inspected in accordance with general
requirements of 1.2.3.4.
3.6 Fungus.
The fungus test is used to determine whether the test item will support fungal
growth and/or whether a fungal environment will adversely affect test item
operation.
3.6.1 General effects. Fungal
growth impairs the functioning or use of equipment by changing the physical
properties of the equipment.
3.6.1.1 Damage mechanisms.
Damage mechanisms are as follows:
3.6.1.1.1 Direct attack on materials.
Non-resistant materials are susceptible to direct attack as the fungi break the
material down and use it as food. This results in deterioration affecting the
physical properties of the material. Examples of non-resistant materials are:
3.6.1.1.1.1 Natural materials.
Products of natural origin are most susceptible to this attack. Examples of
such products are: cellulosic materials (e.g., wood, paper, natural fiber
textiles, and cordage); animal and vegetable based adhesives; grease, oils, and
many hydrocarbons; and leather.
3.6.1.1.1.2 Synthetic materials.
Synthetic products are also susceptible to this attack. Examples of such
products are: polyvinylchloride (PVC) formulations (e.g., those plasticized
with fatty acid esters); certain polyurethanes (e.g., polyesters and some
polyethers); plastics, which contain organic fillers of laminating materials;
and paints and varnishes which contain susceptible constituents.
3.6.1.1.2 Indirect attack on
materials. Damage to fungus resistant materials results from indirect
attack when: fungal growth on surface deposits of dust, grease, perspiration,
and other contaminants find their way onto equipment during manufacture or
accumulate during service causing damage to the underlying material even though
that material may be resistant to direct attack; metabolic waste products
(i.e., organic acids), excreted by fungi, cause corrosion of metals, etching of
glass or staining or degrading of plastics and other materials; and products of
fungal growth on adjacent materials, which are susceptible to direct attack,
come in contact with the resistant materials.
3.6.1.2 Physical interference.
Physical interference can occur as follows:
3.6.1.2.1 Electrical or electronic
systems. Damage to electrical or electronic systems may result from either
direct or indirect attack. Fungal growth can form undesirable electrical
conducting paths across insulating materials or may adversely affect the
electrical characteristics of critically adjusted electronic circuits.
3.6.1.2.2 Optical systems.
Damage to optical systems results primarily from indirect attack. The fungal
growth can adversely affect light transmission through the optical system;
block delicate moving parts; and change non-wetting surfaces to wetting
surfaces with resulting loss in performance.
3.6.1.3 Health and aesthetic factors.
Fungus growth on equipment can cause physiological problems, (e.g., allergies)
or be so aesthetically unpleasant that the users will be reluctant to use the
equipment.
3.6.2 Equipment specification data
required. Before proceeding, verify that the equipment specification
specifies the following as a minimum: pretest data required; failure criteria;
test period if other than 28 days; and whether the equipment should be operated
during test.
3.6.3 Special considerations.
3.6.3.1 Failure analysis. Any
fungi on the test item must be analyzed to determine if the growth is on the test
item material(s) or on contaminants. Any fungal growth on the test item
material(s) (whether from the inoculum or other sources) must be evaluated by
qualified personnel for: the extent of growth on the component(s); the
immediate effect that the growth has on the physical characteristics of the
test item; the long range effect that the growth could have on the test item;
and the specific material(s) (nutrient) supporting the growth. Fungal growth
must not be disturbed during the operational checkout. Human factors effects
must be evaluated.
3.6.3.2 Test interruption. This
policy is designed to provide a standard methodology for selecting a course of
action in the event of an unscheduled interruption in a test. Any deviation
from this policy shall be explained in the test report. Every case of an
interrupted test shall be examined individually using the logical decision
process provided in this document. The fungus test, unlike other environmental
tests, involves living organisms. If the test is interrupted, the fact that
live organisms are involved must be considered.
3.6.3.2.1 Early interruption.
If the interruption occurs early in the test, the test should be restarted from
the beginning with a new test item, or a cleaned item.
3.6.3.2.2 Later interruption.
If the interruption occurs late in the test cycle, examine the item for
evidence of fungal growth. If the test item is biosusceptible, there is no
need for a retest. If there is no evidence of fungal growth or if the
interruption occurred early, proceed as follows:
3.6.3.2.2.1 Lowered temperature.
A lowering of the test chamber temperature generally will retard fungus
growth. If there is no evidence of mycological deterioration and the relative
humidity has been maintained, re-establish test conditions and continue the
test from the point that the temperature fell below the prescribed tolerances.
3.6.3.2.2.2 Elevated temperature.
Elevated temperatures may have a drastic effect on fungus growth. Any rise in
temperature above 88°F for a period of 4 hours or more shall result in complete
re-initiation of the test. If the over temperature condition is less than 4
hours and there is no evidence of mycological deterioration, re-establish test
conditions and continue the test as if no interruption had occurred.
3.6.3.2.2.3 Lowered humidity.
Any drop in humidity levels below 90 percent for a period of 4 hours or more
shall result in complete re-initiation of the test. If the occurrence results
in a more moderate drop in relative humidity that does not result in fungal
deterioration, re-establish test conditions and continue the test as if no
interruption had occurred.
3.6.3.2.3 Reinitiating test.
Although it is preferable to use a new test item when reinitiating a test, the
same test item may be used if it is cleaned. New cotton control strips shall
be placed in the test chamber and both the test item and controls will be
reinoculated with the test fungi.
3.6.4 Apparatus. The required
apparatus consists of chambers or cabinets together with auxiliary
instrumentation capable of maintaining and monitoring the specific conditions
of temperature and humidity and complies with 1.2.4.1.
3.6.5 Preparation for test.
3.6.5.1 Chamber and accessories.
The chamber and accessories shall be constructed and arranged in such a manner
as to avoid condensate dripping on the test item. The chamber shall be vented
to the atmosphere to prevent the buildup of total pressure.
3.6.5.2 Relative humidity (RH).
Relative humidity shall be determined from the dry bulb - wet bulb temperature
comparison method or an equivalent method approved by the procuring activity.
When the wet bulb control method is used, the wet bulb assembly shall be
cleaned and a new wick installed for each test. The air velocity across the
wet bulb shall not be less than 900 feet per minute. The wet and dry bulb
sensors shall not be installed in the discharge side of any local fan or blower
used to create the requirements of 3.6.5.3.
3.6.5.3 Air flow. Provisions
shall be made for controlling the flow of air throughout the internal test
chamber space so that the air velocity shall be between 197-394 ft/min. Free
circulation of air around the test item shall be maintained, and the contact
area of fixtures supporting the test item shall be kept to a minimum (see
1.2.4).
3.6.5.4 Data to be recorded.
Continuous recordings of test section temperature and relative humidity
conditions shall be taken. Readout charts shall be capable of being read with
a resolution within 1°F. The desired humidity shall be generated by using
steam or distilled, demineralized or deonized water having a pH value between
6.0 and 7.2 at 73°F.
a. Live
steam shall not be injected directly into the test section working space where
it may have an adverse affect on the test item and microbial activity.
b. Rust
or corrosive contaminants shall not be imposed on the test item by the test
facility.
3.6.5.5 General conditions.
Unless otherwise specified: all reagents shall conform to the specification of
the Committee on Analytical Reagents of the American Chemical Society, where
such specifications are available; and references to water shall be understood
to mean distilled water or water of equal purity. If the test is interrupted,
follow the procedure of 3.6.3.2.
3.6.6 Procedures. The following
test phases are provided for application in combinations, as required.
3.6.6.1 Phase I, pre-test checkout.
All items require a pre-test checkout to provide baseline data. Conduct the checkout
as follows:
a. Prepare
the test item in accordance with 1.2.3.2 and required test item configuration.
b. Conduct
a complete visual examination of the test item with special attention to
discolored areas, imperfections, or the existence of any other conditions that
could be conducive to fungus growth.
c. Document
the results of step b.
d. Conduct
an operational checkout in accordance with the approved test plan if operation
is specified by the requirements document during or following the fungus test.
e. Record
results for compliance with 1.2.3.1.
f. Proceed
to phase II.
3.6.6.2 Phase II, chamber test.
3.6.6.2.1 Preparation.
3.6.6.2.1.1 Preparation of
mineral-salts solution.
a. Using
clean apparatus, prepare the mineral salts solution to contain the following:
Potassium dihydrogen
orthophosphate (KH2PO4)
0.7 g
Potassium monohydrogen
orthophosphate (K2HPO4) 0.7
g
Magnesium sulfate heptahydrate
(MgSO4·7H2O) 0.7
g
Ammonium nitrate (NH4NO3) 1.0
g
Sodium chloride (NaC1) 0.005
g
Ferrous sulfate heptahydrate
(FeSO4·7H2O) 0.002 g
Zinc sulfate heptahydrate
(ZnSO4·7H2O) 0.002
g
Manganous sulfate monohydrate
(MnSO4·H2O) 0.001 g
Distilled water 1000
ml
b. Measure
the pH of the mineral salts solution. If not between 6.0 and 6.5, discard and
prepare a proper solution.
3.6.6.2.1.2 Preparation of mixed
spore suspension. NOTE: PRECAUTIONS: Although the fungi
specified for this test are not normally considered to present a serious hazard
to humans, certain people may develop allergies or other reactions. It is
therefore recommended that standard operation procedures (SOPs) for safety be
employed. It is also recommended that the tests be conducted by personnel
trained in microbiological techniques.
a. Using
aseptic techniques, prepare the spore suspension containing the test fungi
specified in Table I.
TABLE I. Test fungi.
Fungus
source identification no.
Fungi 1/ USDA
2/ ATCC 3/
Aspergillus niger QM
386 ATCC 9642
Aspergillus flavus QM
380 ATCC 9643
Aspergillus versicolor QM
432 ATCC 11730
Penicillium funiculosum QM
474 ATCC 11797
Chaetomium globosum QM
459 ATCC 6205
Notes:
1. The
fungi may be distributed in a lyophilized state, or on agar slants.
2. U.S.
Department of Agriculture (SEA/FR), Northern Regional Research Center, ARS
Culture Collection, 1815 N. University Street, Peoria, Illinois 60604
3. American
Type Culture Collection, 12301 Parklawn Drive,
Rockville, Maryland
20852
b. Maintain
pure cultures of these fungi separately on an appropriate medium such as potato
dextrose agar except that chaetomium globosum shall be cultured on strips of
filter paper overlayed on the surface of mineral salts agar.
c. Prepare
mineral salts agar by dissolving 15.0g agar in a liter of the mineral salts
solution described in 3.6.6.2.1.1. Note: Do not keep the stock
cultures for more than 4 months at 43 ± 7°F; after that time, prepare
subcultures and use them for the new stocks.
d. Verify
the purity of fungus cultures prior to the test.
e. Incubate
subcultures used for preparing new stock cultures or the spore suspension at
86.0 ± 2.5°F for 14 to 21 days.
f. Prepare
a spore suspension of each of the five fungi by pouring into one subculture of
each fungus 10 ml of an aqueous solution containing 0.05g per liter of a
nontoxic wetting agent such as sodium dioctyl sulfosuccinate or sodium lauryl
sulfate.
g. Use
a rounded glass rod to gently scrape the surface growth from the culture of the
test organism.
h. Pour
the spore charge into a 125 ml capped Erlenmeyer flask containing 45 ml of
water and 50 to 75 solid glass beads, 5 mm in diameter.
i. Shake
the flask vigorously to liberate the spores from the fruiting bodies and to
break the spore clumps.
j. Filter
the dispersed fungal spore suspension into a flask through a 6 mm layer of
glass wool contained in a glass funnel. Note: This process should
remove large mycelial fragments and clumps of agar.
k. Centrifuge
the filtered spore suspension and discard the supernatant liquid.
l. Resuspend
the residue in 50 ml of water and centrifuge. Wash the spores obtained from
each of the fungi in this manner three times.
m. Dilute
the final washed residue with mineral-salts solution in such a manner that the
resultant spore suspension shall contain 1,000,000 ± 200,000 spores per ml as
determined with a counting chamber.
n. Repeat
this operation for each organism used in the test.
o. Perform
a viability check for each organism in accordance with 3.6.6.2.1.3.
p. Blend
equal volumes of the resultant spore suspension to obtain the final mixed spore
suspension. Note: The spore suspension may be prepared fresh. If not
freshly prepared, it should be held at 43 ± 7°F for not more than 7 days.
3.6.6.2.1.3 Control items. Two
types of control tests are required. Using the procedure of 3.6.6.2.1.3.1,
verify the viability of the spore suspension and its preparation. Using the
procedure of 3.6.6.2.1.3.2, verify the suitability of the chamber environment.
3.6.6.2.1.3.1 Viability of spore
suspension. Prior to preparing the composite spore suspension, inoculate
sterile potato dextrose agar plates with 0.2 to 0.3 milliliters of the spore
suspension of each of the individual fungal species using separate potato
dextrose agar plates for each species.
a. Distribute
the inoculum over the entire surface of the plate.
b. Incubate
the inoculated potato dextrose agar plate at 75 to 88°F for 7 to 10 days.
c. After
the incubation period, check the fungal growth. Note: The absence of
copious growth of any of the test organisms over the entire surface in each
container will invalidate the results of any tests using these spores.
3.6.6.2.1.3.2 Test chamber
environment. Inoculate a known susceptible substrate along with the test
item to ensure that proper conditions are present in the incubation chamber to
promote fungal growth. Note: The controlled substrate shall consist of
cotton fabric strips conforming to MIL-T-43566A Tape, Type 1a,
Class 2, bleached, white flat construction.
a. Prepare
the following solution:
(1) 10.0 grams glycerol
(2) 0.1 potassium
dihydrogen orthophosphate (KH2PO4)
(3) 0.1g ammonium nitrate
(NH4NO3)
(4) 0.025g magnesium
sulfate heptahydrate (MgSO4·7H2)
(5) 0.05g yeast extract
(6) Distilled water to a
total volume of 100 ml
(7) Hydrochloric
acid (HC1) and sodium hydroxide (NaOH) base, as required, to adjust the final
solution pH to 5.3.
b. Dip
the cotton strips into the above solution. After dipping, remove the excess
liquid from the strips and hang them to dry before placing them in the chamber
and inoculating them.
c. Within
the chamber, place the strips vertically in proximity to and bracketing the
test item so that the test strips and test items experience the same test
environment. The length of the strips shall be at least the height of the test
item.
3.6.6.3 Phase III, test performance.
3.6.6.3.1 Preparation for incubation.
Assure that the condition of the test items subjected to testing is similar to
that as delivered by the manufacturer or customer for use, or as otherwise
specified. Any cleaning of the test item shall be accomplished at least 72
hours prior to the beginning of the fungus test.
a. Install
the test item in the chamber or cabinet on suitable fixtures or suspended from
hangers.
b. Hold
the test item in the operating chamber for at least 4 hours immediately prior
to inoculation.
c. Inoculate
the test item and cotton fabric chamber control items (3.6.6.2.1.3.2) with the
mixed fungal spore suspension (3.6.6.2.1.2) by spraying it on the control and
on and into the test item(s) (if not hermetically sealed) 1/ in the form
of a fine mist from an atomizer or nebulizer. Note: In spraying the
test and control items with composite spore suspension, care should be taken to
cover all external and internal surfaces which are exposed during use or
maintenance. If the surfaces are nonwetting, spray until initiation of droplet
coalescence.
d. Replace
covers of the test items loosely.
e. Start
incubation immediately following the inoculation.
1/ Personnel with appropriate knowledge of
the test item should be available to aid in exposing its interior surfaces for
inoculation.
3.6.6.3.2 Incubation of the test
item. Incubate the test item(s) under a daily cycle of temperature and
humidity conditions consisting of 20 hours at a relative humidity of 95 ± 5
percent at an air temperature of 86 ± 2°F followed by a 4-hour period in which
conditions of 95 percent relative humidity at 77 ±2°F are maintained for at
least 2 hours. Up to a total of 2 hours of the 4-hour period will be used for
the transition(s) of temperature and relative humidity. Temperature and
humidity conditions during the transition periods shall be controlled as
follows: temperature 75° to 88°F and relative humidity above 90 percent.
a. Repeat
the 24-hour daily cycle for the test duration.
b. After
7 days, inspect the growth on the control cotton strips to assure that the
environmental conditions in the chamber are suitable for growth. For this
assurance, at least 90 percent of the part of the surface area of each test
strip located at the level of the test item should be covered by fungi when
inspected visually. If not, repeat the entire test with the required
adjustments of the chamber to produce conditions suitable for growth. Leave
the control strips in the chamber for the duration of the test; note their
condition at this time and record it with the test item data as described in
3.6.6.1, step c.
c. If
the cotton strips show satisfactory fungus growth after 7 days, continue the
test for the required period of days from the time of inoculation. If there is
a decrease in fungal growth on the cotton strips at the end of the test as
compared to the 7-day results, the test is invalid.
3.6.6.3.3 Inspection. At the
end of the incubation period, inspect the test item immediately. If possible,
inspect the item within the chamber. If the inspection is conducted outside of
the chamber and not completed in 8 hours, return the test item to the test
chamber or similar humid environment for a minimum of 12 hours. Except for
hermetically sealed equipment, open the equipment enclosure and examine both
the interior and exterior of the test item. Record results of the inspection
to include applicable information. Note: Data shall be used for
comparison with that obtained in 3.6.6.1 c.
3.6.6.4 Phase III, operation/usage.
If operation is required (e.g. electrical equipment), conduct the operation in
the period as specified in 3.6.6.3.3. Data shall be recorded for comparison
with the baseline data obtained in 3.6.6.1 c. Personnel with appropriate
knowledge should be available to aid in exposing interior surfaces of the item
for inspection and making operational and use decisions.
3.7 Dust and
sand. The dust and fine sand test is used to ascertain the ability of
equipment to resist the effects of a dry dust and sand laden atmosphere. This
test simulates the effect of sharp edged dust and fine sand particles, up to
150 microns in size or as otherwise specified in the equipment specification,
that may penetrate into cracks, crevices, bearings, and joints. This test is
applicable to all mechanical, electrical, electronic, electrochemical, and
electromechanical devices for which exposure to the effects of a dry dust or
fine sand laden atmosphere is anticipated.
3.7.1 General effects. General
effects resulting from the penetration of dust and fine sand, hereafter also
referred to as contaminants,, can cause a variety of damage such as fouling
moving parts, making relays inoperative, forming electrically conductive
bridges with resulting shorts and acting as a nucleus for the collection of
water vapor, and hence a source of possible corrosion and malfunction of
equipment.
3.7.2 Equipment specification data
required. Before proceeding, verify that the equipment specification
specifies the following as a minimum: pretest data required; failure criteria;
change in orientation during test, if required; whether the equipment is
supposed to operate during test and for how long; and whether the equipment
shall be subjected to sand, dust, or sand and dust environments.
3.7.3 Apparatus.
3.7.3.1 Test facility. The test
facility shall consist of a chamber (see 1.2.4.1) with accessories to control
dust concentration, velocity, temperature, and humidity of dust laden air. In
order to provide adequate circulation of the dust and sand laden air, no more
than 50 percent of the cross-sectional area (normal to air flow) and 30 percent
of the volume of the chamber shall be occupied by the test item(s). The
chamber shall be provided with a suitable means of maintaining and verifying
the dust and sand concentration in circulation. A minimum acceptable means for
doing this is by use of a properly calibrated smoke meter and standard light
source. The dust and sand laden air shall be introduced into the test space in
such a manner as to allow it to become approximately laminar in flow before it
strikes the test item.
3.7.3.2 Dust and sand
characteristics. Unless otherwise specified in the equipment
specification, contaminants used in this test shall be dry dust and a fine
sand, both 97-99 percent by weight silicon dioxide (Si02), of
angular structure, and shall have the size distribution determined by weight,
as specified in Table II. The typical composition of the dust and quartz sand
used shall be as specified in Table III.
3.7.4 Procedures. Prepare the
test item in accordance with the general requirements of 1.2.3 positioning the
test item as near the center of the chamber as practical. If more than one
item is being tested, there shall be a minimum clearance of 4 inches between
surfaces of test items or any other material or object capable of furnishing
protection. Also, no surface of the test item shall be closer than 4 inches
from any wall of the test chamber. Orient the item so as to expose the most
critical or vulnerable parts to the dust stream. Proceed as follows unless
otherwise specified in the equipment specification and depending on whether the
dust and sand tests are to be performed separately or together as deemed
acceptable by the procuring activity.
3.7.4.1 Dust.
a. Set
the chamber controls to maintain an internal chamber temperature of 73°F and a
relative humidity of less that 22 percent. Adjust the air velocity to 1,750
±250 feet per minute (fpm). Adjust the dust feeder to control the dust
concentration at 0.3 ±0.2 grams per cubic foot. With the test item nonoperating,
maintain these conditions for 6 hours.
b. Stop
the dust feed and reduce the air velocity to 300 ±200 fpm. Raise the internal
chamber air temperature to 145°F. Hold these conditions 16 hours.
c. While
holding chamber temperature at 145°F, adjust the air velocity to 1,750 ±250
fpm. Adjust the dust feeder to control the dust concentration at 0.3 ±0.2
grams per cubic foot. Unless otherwise specified, with the test item
nonoperating, maintain these conditions for 6 hours.
d.
Turn off all chamber controls
and allow the test item to return to standard ambient conditions. Remove
accumulated dust from the test item by brushing, wiping, or shaking while
exercising care to avoid introduction of additional dust into the test item.
Dust shall not be removed by either air blast or vacuum cleaning.
e. Operate
and inspect the test item in accordance with 1.2.3.4.
f. Inspect
the test item and obtain results as specified in the general requirements of
1.2.3.4. In the performance of this inspection, test items containing
bearings, grease seals, lubricants, et cetera, shall be carefully examined for
the presence of dust deposits.
TABLE II. Dust and sand
particle size blend.
Particle Blend
by weight (percent)
size range
(Microns) Sand
blend Dust blend
1000.0 - 400.0
14.3 0.0
400.0 - 200.0 18.7 0.0
200.0 - 80.0 31.1 6.3
80.0 - 40.0 10.6 21.5
40.0 - 20.0 6.6 18.9
20.0 - 10.0 7.2 20.5
10.0 - 5.0 6.4 18.2
5.0 - 2.5 2.5 7.3
2.5 - 1.0 1.2 3.3
1.0 - 0.1 1.4 4.0
0.1 - 0.0 - -
Note: Mixture conforms
to SAE J726.
TABLE III. Dust/quartz sand
composition.
Composition
by weight
Material (percent)
Silicon
dioxide 99.500
Rare earth
oxides (R2O3) 0.200<
Chromium 0.002<
Phosphorus 0.010<
Carbonate 0.010<
Trapped
water 0.500<
Note:
Mixture conforms to SAE J726.
3.7.4.2 Sand.
a. Set
the chamber controls to maintain an internal chamber temperature of 73°F and a
relative humidity of less that 22 percent. Adjust the air velocity to 1,750
±250 fpm. Adjust the sand feeder to control the sand concentration at 0.3 ±0.2
grams per cubic foot. With the test item nonoperating, maintain these
conditions for 0.5 hour.
b. Stop
the sand feed and reduce the air velocity to 300 ±200 fpm. Raise the internal
chamber air temperature to 145°F. Hold these conditions 16 hours.
c. While
holding chamber temperature at 145°F adjust the air velocity to 1,750 ±250
fpm. Adjust the sand feeder to control the sand concentration at 0.3 ±0.2
grams per cubic foot. Unless otherwise specified, with the test item
nonoperating, maintain these conditions for 0.5 hour.
d. Turn
off all chamber controls and allow the test item to return to standard ambient
conditions. Remove accumulated sand from the test item by brushing, wiping, or
shaking while exercising care to avoid introduction of additional sand into the
test item. Sand shall not be removed by either air blast or vacuum cleaning.
e. Operate
and inspect the test item in accordance with 1.2.3.4.
f. Inspect
the test item and obtain results as specified in the general requirements of
1.2.3.4. In the performance of this inspection, test items containing
bearings, grease seals, lubricants, et cetera, shall be carefully examined for
the presence of sand deposits.
3.7.4.3. Dust and sand combined.
a. Set
the chamber controls to maintain an internal chamber temperature of 73°F and a
relative humidity of less that 22 percent. Adjust the air velocity to 1,750
±250 fpm. Adjust the dust feeder to control the dust concentration at 0.3 ±0.2
grams per cubic foot. With the test item nonoperating, maintain these
conditions for 5.5 hours.
b. While
maintaining the conditions of step a, adjust the sand feeder to control the
sand concentration at 0.3 ±0.2 grams per cubic foot and continue testing for an
additional 0.5 hour.
c. Stop
the dust and sand feed and reduce the air velocity to 300 ±200 fpm. Raise the
internal chamber air temperature to 145°F. Hold these conditions 16 hours.
d. While
holding chamber temperature at 145°F, adjust the air velocity to 1,750 ±250
fpm. Adjust the dust feeder to control the dust concentration at 0.3 ±0.2
grams per cubic foot. Unless otherwise specified, with the test item
nonoperating, maintain these conditions for 5.5 hours.
e. While
maintaining the conditions of step d, adjust the sand feeder to control the
sand concentration at 0.3 ±0.2 grams per cubic foot and continue testing for an
additional 0.5 hour.
f. Turn
off all chamber controls and allow the test item to return to standard ambient
conditions. Remove accumulated dust and sand from the test item by brushing,
wiping, or shaking while exercising care to avoid introduction of additional
dust into the test item. Dust shall not be removed by either air blast or
vacuum cleaning.
g. Operate
and inspect the test item in accordance with 1.2.3.4.
h. Inspect
the test item and obtain results as specified in the general requirements of
1.2.3.4. In the performance of this inspection, test items containing
bearings, grease seals, lubricants, et cetera, shall be carefully examined for
the presence of dust and sand deposits.
3.8 Rain and
wind. The rain and wind test is conducted to determine the
effectiveness of protective covers or cases to shield equipment from rain.
This test is applicable to equipment which may be exposed to rain under service
conditions.
3.8.1 Equipment specification data
required. Before proceding, verify that the equipment specification
specifies the following as a minimum: pretest data required; failure criteria;
and whether the equipment is to operate for a specified period of time during
the rain fall.
3.8.2 Apparatus. A chamber
having the general characteristics of 1.2.4.1, but with rain and wind source,
is required. The rain chamber shall have the capability of producing falling
rain and wind blowing horizontally at up to 40 miles per hour (35 knots). The
chamber temperature shall be uncontrolled, except as regulated by water
introduced as rain, throughout the test period.
a. The
rain shall be produced by a water distribution device of such design that the
water is emitted in the form of droplets having a diameter range between 0.5
and 4.5 millimeters. (NOTE: A recommended method of measuring raindrop size is
the flour pellet method as referenced in "The Relation of Raindrop Size to
Intensity" by Laws and Parsons, Transactions of the American Geophysical
Union, Part II, pps 452 to 459, -1943.) The temperature of the water shall be
between 52° and 95°F. The water distribution shall be such that, with the wind
source turned off, the rain is dispersed completely over the test item; an
additional test area shall be provided to accomplish the wind blown rain
effects on the equipment.
b.
The wind source shall be
positioned with respect to the test item such that it will cause the rain to
beat directly (with variations up to 45 degrees from the horizontal) and
uniformly against one side of the test item. The wind velocity shall be
measured at the position of the test item, prior to placement of the item in
the chamber. No rust or corrosive contaminants shall be imposed on the test
item by the test facility.
3.8.3 Procedure. The procedure
is intended for evaluating equipment used outdoors under blowing rain
conditions. All test conditions shall apply except as otherwise stated in the
equipment specification. The test item shall be placed in the chamber in its
normal operating position in accordance with 1.2.3.2. The test item shall be
exposed to a simulated rain at a rate of 2 + 0.5 inches per hour for 10
minutes. The rate of rainfall shall then be raised to 5 + 1 inches per
hour and held at this rate for 5 minutes. The rate shall then be reduced to 2 +
0.5 inches per hour for the next 15 minutes. Starting 5 minutes after the
initiation of the rain, the wind source shall be turned on and adjusted to
produce a horizontal wind velocity of 40 miles per hour (3,500 feet per
minute). The wind source shall be maintained at this velocity for 15 minutes
after which the wind source shall be turned off.
NOTE: If
specified in the equipment specification, the test item shall be operated
during the last 10 minutes of the 30-minute rain. Each of the sides of the
test item that could be exposed to blown rain shall be subjected to the rain
for a period of not less than 30 minutes, for a total test duration of not less
than 2 hours. At the conclusion of the test period, the test item shall be
removed from the test chamber, operated, inspected, and the results obtained
shall be recorded in accordance with 1.2.3.4. The protective cover or case
shall, where possible, then be removed and the test item inspected for
compliance with 1.2.3.4.
3.9 Solar
radiation. This test is conducted to determine the effects of solar
radiation on equipment that may be exposed to sunshine during operation or
unsheltered storage on the earth's surface or in the lower atmosphere.
3.9.1 General effects.
3.9.1.1 Heating. The heating
effects of solar radiation differ from those of high air temperature alone in
that the amount of heat absorbed or reflected depends on the roughness and
color of the surface on which the radiation is incident. In addition to the
differential expansion between dissimilar materials, changes in the intensity
of solar radiation may cause components to expand or contract at different
rates, which can lead to severe stresses and loss of structural integrity. In
addition to those specified in 3.1.1.1, some other examples of heating effects
include:
a. Jamming
or loosening of moving parts.
b. Weakening
of solder joints and glued parts.
c. Change
in strength and elasticity.
d. Loss
of calibration or malfunction of linkage devices.
e. Loss
of seal integrity.
f. Changes
in electrical or electronic components.
g. Premature
actuation of electrical contacts.
h. Fading
of colors of color-coded components.
i. Changes
in characteristics of elastomers and polymers.
j. Blistering
and peeling of paints and other finishes.
k.
Softening of potting compounds.
3.9.1.2 Actinic. In addition to
the heating effects of 3.9.1.1, which are caused by the infrared portion of the
solar spectrum, certain degradation from solar energy may be attributable to
other portions of the spectrum, particularly the ultraviolet. Since the rate
at which these reactions will occur generally increases as the temperature
rises, the full spectrum must be used to adequately simulate the actinic
effects of solar radiation. Some examples of deterioration caused by actinic
effects are:
a. Fading
of fabric color.
b. Checking
and fading of paints.
c. Deterioration
of natural and synthetic elastomers and polymers through photochemical
reactions initiated by shorter wavelengths radiation.
3.9.2 Equipment specification data
required. Before proceeding, verify that the equipment specification
specifies the following as a minimum: test item configuration and orientation;
pretest data required; failure criteria; number of test cycles and temperature
and solar radiation intensity if other than that given in the following
procedure; and whether the equipment is to operate for a specified period of
time during exposure. Consideration shall be given to using this method of
test as an alternative to performing the high temperature test of 3.1.1.
3.9.3 Guidelines for test conditions.
3.9.3.1 Application. Apply this
test method when the test item is likely to be exposed to solar radiation in
the open in hot climates during its life cycle, and when the effects mentioned
in 3.9.1 are of concern. In most cases, this method should replace the high
temperature test of 3.1.1.
3.9.3.2 Restrictions. This test
method is not to be used to simulate the heating effects that result from
enclosed or covered storage conditions. The solar spectrum and energy levels
are those that are received at sea level. The ultraviolet portion is simulated
only in a general way, but is considered adequately representative of levels in
most geographic areas.
3.9.3.3 Configuration. The test
item configuration should be the same as its configuration during exposure to
solar radiation. The orientation of the test item relative to the direction of
radiation will have a significant impact on the heating effects, as will its
mounting (on supports or on a substrate of specified properties, e.g., a layer
of concrete of specified thickness or a sand bed of certain reflectivity).
3.9.4 Apparatus.
3.9.4.1 Test facility. The
required facility consists of a chamber or cabinet, auxiliary instrumentation,
and a solar lamp bank. This apparatus must be capable of maintaining and
monitoring the required conditions of temperature, airflow, and irradiation.
3.9.4.1.1 Test chamber. The
volume of the test chamber shall be a minimum of 10 times that of the envelope
volume of the test item.
3.9.4.1.2 Solar lamp bank. The
solar radiation source area shall be such that the length and width of the test
item shall be no more than one-half the same dimensions of the lamp bank and
shall be composed of lamps that simulate the solar spectrum.
a. The
irradiance shall have a maximum intensity of 1120 Watts per meter squared (W/m2)(±10%),
and the radiation falling on the test item shall be uniform to within ±10
percent of the desired value, with the spectral distribution given in Table
IV. Where thermal effects only are to be assessed, deviation from this
spectral distribution is permitted, but the irradiance must be adjusted to give
an equivalent heating effect. In order to calculate this adjustment, it is
necessary to know: the spectral reflectance or transmittance of the irradiated
surfaces; and the spectral energy distribution of the particular lamps being
used (and also the effect of any associated reflectors or glasses).
b. The
radiation shall be directed onto the test item and shall irradiate the entire
surface of the test item facing the solar radiation source. This value of 1120
W/m2 shall include any radiation reflected from the test chamber
walls and received by the test item, but it should not include long-wave
infrared radiation emitted by the chamber walls. The radiation-measuring
device shall be calibrated in the wavelength range of the test source
radiations.
c. The
radiation source shall be located at least 30 inches away from any other
surface of the test item.
3.9.4.1.3 Light source. Tests
conducted for degradation and deterioration of materials due to actinic
effects, as well as heat buildup within the test items, must satisfy the full
spectrum of Table IV and may use one of the following acceptable radiation
sources:
TABLE IV. Spectral energy
distribution and permitted tolerances.
Spectral
region
Ultraviolet
Characteristic Min.
Max. Visible Infrared
Bandwidth 0.28 to 0.32
to 0.40 to 0.78 mm 0.78 to 3.00 mm
0.32 0.40
µm
micrometer
(µm)
Irradiance 5 W/m2 63
W/m2 517 to 604 W/m2 492 W/m2
Tolerance ±35% ±25% ±10%
±20%
NOTE: The
amount of radiation wavelength shorter than 0.30 µm reaching Earth's surface is
insignificant
a. Xenon
arc or mercury xenon arc (used singly) lamps with suitable reflector.
b. Combination
of high pressure sodium vapor and improved mercury vapor lamps with suitable
reflectors.
c. High-intensity
multivapor, mercury vapor (with suitable reflectors), and incandescent spot
lamps.
d. Carbon arc lamps with
suitable reflectors.
NOTE: Other
combinations of the lamps listed above and in e through k below may be used if
it is proven that the combination produces the spectrum of Table IV.
e. Mercury vapor lamps
(internal reflector type only).
f. Combination
of incandescent spot lamps and tubular-type mercury vapor lamps with external
reflectors.
g. Combination
of incandescent spot lamps and mercury vapor lamps with internal reflectors.
h. Metal halide lamps.
i. Mercury xenon arc lamps
with suitable reflectors.
j. Multivapor lamps (clear or
coated bulb) with suitable reflectors.
k. Tungsten filament lamps.
The preceding list is not intended to exclude new
lamps made available by advanced technology.
3.9.4.2 Controls.
3.9.4.2.1 Temperature. Chamber
air temperature shall be maintained in accordance with the general requirements
of 1.2.1 and measured (with adequate shielding from radiated heat) at a point
or points in a horizontal plane 0 to 2 inches below the prescribed irradiation
plane, at half the distance between the test item and the wall of the chamber
or at 40 inches from the test item, whichever is smaller. This is to ensure
reasonable control of the envelope of air surrounding the test item.
3.9.4.2.2 Surface contamination.
Dust and other surface contamination may significantly change the absorption
characteristics of irradiated surfaces. Unless otherwise required, specimens
should be clean when they are tested. However, if effects of surface
contamination are to be assessed, the relevant specification should include the
necessary information on preparation of surfaces.
3.9.4.2.3 Instrumentation.
ITEM TOLERANCE
Pyranometer
or pyrheliometer Total irradiation (direct
and scattered) to ±47 W/m2 (±14 British thermal units per foot
squared per hour (Btu/ft2/hr))
Spectroradiometer
or filtered ±5% of reading.
pyranometer
NOTE: Values may
be assumed to represent plus or minus two standard deviation; thus, the stated
tolerances should not be exceeded in more than 1 measurement out of 20. Solar
radiation intensity shall be measured with a pyranometer or pyrheliometer.
Spectral distribution of irradiance as a function of wavelength shall be
measured with a spectral radiometer or filtered pyranometer.
3.9.4.2.4 Calibration of chamber.
Because of the variety of permissible lamps and chamber designs, it is
particularly important that the chamber be calibrated to assure that the proper
levels of radiant infrared energy are impacting the test area when heat alone
is of concern and that the proper intensity and spectral distribution of solar
radiation are impacting the test area when actinic effects are of concern.
Over the area covered by the test item, the radiation intensity must be within
±10 percent. As the lamps age, their spectral output changes. To ensure that
solar radiation chambers meet established specifications, a check on spectral
distribution, intensity, and uniformity shall be performed at intervals not
exceeding 500 hours of operation to ensure that the facilities continue to meet
established specifications. This value is based on the manufacturer's
guarantee for minimum bulb life.
3.9.5 Preparation for test.
3.9.5.1 Preliminary steps.
Before initiating any testing, determine from the test plan:
a. Which
test procedures are required.
b. The
diurnal cycle to be used.
c. Other variables, such as number
of cycles, etc.
3.9.5.2 Pretest standard ambient
checkout. All items require a pretest standard ambient checkout to provide
baseline data. Conduct the checkout as follows:
a. Install
the test item in the chamber and stabilize it at standard ambient conditions
(general requirements, 1.2.1) and in a manner that will simulate service usage,
unless the storage configuration is specified. Position the test item in
accordance with the following:
(1) As
near the center of the test chamber as practical and so that the surface of the
item is not closer than 1 ft to any wall or 30 in. to the radiation source when
the source is adjusted to the closest position it will assume during the test.
(2) Oriented,
within realistic limits, to expose its most vulnerable parts to the solar
radiation, unless a prescribed orientation sequence is to be followed.
(3) Separated
from other items that are being tested simultaneously, to ensure that there is
no mutual shading or blocking of airflow.
b. Conduct
a visual examination of the test item with special attention to stress areas,
such as corners of molded cases and document the results.
c. Prepare
the test item in accordance with the test item configuration of 3.9.3.4 and
with the temperature sensors necessary to determine test item response.
d. Conduct
an operational checkout in accordance with the approved test plan and record
results.
e. If
the test item operates satisfactorily, place it in its test configuration (if
other than operational). If not, resolve the problem and restart at step a.
Position the test item in accordance with step a preceding and proceed to the
first test as specified in the test plan.
3.9.6 Procedures. The following
test procedure provides the basis for evaluating the performance of the test
item in a solar radiation environment. Cycling and solar radiation shall be in
accordance with Figure 9.
a. Adjust
the solar radiation source to a radiant energy rate of 1120 + 47 W/m2
(355 + 14 Btu/ft2/hr), or as specified in the equipment
specification, and a temperature of 120°F.
b. Maintain
these conditions for 20 hours. If required, operational checks should be
conducted during the last 4 hours of each 20-hour exposure when test
temperatures are maximized.
c. Measure
and record the test item temperatures.
d. Turn
off the solar radiation source for 4 hours.
e. Repeat
steps a thru d for the number of cycles specified in the equipment
specification.
f. Conduct
a performance check and visual examination as in 3.9.5.2 steps b and d and
document the results.
3.10 Cleaning
spray. The cleaning spray test is performed to determine the
resistance of equipment to cleaning spray that will be encountered during
vehicle cleaning under storage and service conditions.
3.10.1 General effects. The
force combined with the distance of application of cleaning spray may cause
physical damage to the part or to the packing or gaskets of the part. Damage
to the packings or gaskets integral to the part or used for interfacing the
part with the vehicle could cause leaks that would be detrimental to personnel
and hardware operation under adverse environmental conditions.
3.10.2 Equipment specification data
required. Before proceeding, verify that the equipment specification
specifies the following as a minimum: required pretest data; failure criteria;
surface area coverage rate, nozzle size, pressure at the nozzle and cleaning
media; and distance of application of the spray from the equipment surface.
3.10.3 Procedures. Proceed as
follows depending on whether the equipment under test is to be mounted external
or internal to the vehicle.
NOTE: For
equipment mounted both external and internal to a vehicle (e.g., a periscope),
the procedure applying to the respective portions shall be followed.
3.10.3.1 Externally mounted
equipment. Mount the assembly so that each of the surfaces that normally
could be subjected to the full force of the spray are accessible. Then,
perform surface cleaning at a rate of 1.0 square foot per minute for a minimum
of 10 minutes (or as otherwise specified in the equipment specification)
followed by a cold water jet rinse of equal duration as follows: apply a steam
and water jet spray with cleaning agents conforming to P-C-437
perpendicular to and at a distance of 1.0 + 0.5 foot from the surfaces
being cleaned; then, using a nozzle having an orifice diameter of 0.25 inch and
a nozzle pressure of 110 psig, apply a cold water jet spray perpendicular to
and at a distance of 5.0 + 0.5 feet from the surfaces being rinsed.
Following test, operate and inspect the test item and obtain results in
accordance with 1.2.3.4.
3.10.3.2 Internally mounted
equipment. Mount the assembly so that each of the surfaces that normally
could be subjected to the full force of the spray are accessible. Using a
nozzle having an orifice diameter of 0.25 inch and a nozzle pressure of 25
psig, apply a cold water jet spray perpendicular to and at a distance of 1.0 +
0.5 foot from the surface being sprayed; clean at the rate of 1.0 square foot
per minute for a minimum of 10 minutes (or as otherwise specified in the
equipment specification). Following test, operate and inspect the test item
and obtain results in accordance with 1.2.3.4.

3.11 Submergence.
The submergence test is performed to determine the adequacy of the equipment
seals to protect the equipment interior from contaminants that will be
encountered during storage and service conditions.
3.11.1 General effects.
Improper sealing of equipment could allow the entrance of contaminants into the
assembly that will adversely affect equipment performance. E.g., moisture or dust
particles within an optics device could obscure the users vision; moisture
entering an electronics unit could cause component shorting; etc.
3.11.2 Equipment specification data
required. Before proceeding, verify that the equipment specification specifies
the following as a minimum: required pretest data; failure criteria;
differential pressure and depth to which the unit is to be submerged, and
duration of test.
3.11.3 Apparatus. A vessel
capable of holding clean water of sufficient depth to allow covering the
assembly completely as specified in the equipment specification is required.
In the case of the procedure of 3.11.4.2, the vessel shall be capable of
introducing and sustaining a differential pressure, in the form of a vacuum, as
specified in the equipment specification.
3.11.4 Procedures. Proceed as
follows depending on whether the equipment under test is equipped with or
without air plugs or valves.
3.11.4.1 Equipment with air plugs or
valves. Remove the air plug from the assembly and install a suitable
pressure adapter into the open port; if there is a Schrader air valve, remove
the valve cap. Install mating connectors on all non-hermetically sealed
electrical connectors and pressurize the assembly to the level specified in the
equipment specification. Totally submerge the assembly in clear water to the
depth specified in the equipment specification (submergence depth shall be
measured between the water surface and top surface of the assembly) and monitor
the assembly for a minimum of 5 minutes (or as otherwise specified in the
equipment specification) for leakage as evidenced by any air bubbles escaping
from the interior of the assembly (be careful not to misconstrue bubbles from
the assembly surface as air escaping from the unit). Any evidence of leakage
shall be cause for rejection. Following the test interval, remove the assembly
from the water, dry the assembly exterior surfaces, vent the assembly interior
to atmospheric pressure, reinstall the air plug or valve cap, and operate and
inspect the test item obtaining results in accordance with 1.2.3.4.
3.11.4.2 Equipment without air plugs
or valves. Install mating connectors on all non hermetically sealed
electrical connectors. Place the assembly into a vessel that is filled with
clear water and that is capable of introducing and sustaining a differential
pressure (vacuum) as specified in the equipment specification. Submerge the
assembly to the depth specified in the equipment specification (submergence
depth shall be measured between the water surface and top surface of the
assembly) and apply the differential pressure for a minimum of five minutes.
Monitor the assembly for leakage as evidenced by any air bubbles escaping from
the interior of the assembly (be careful not to misconstrue bubbles from the
assembly surface as air escaping from the unit). Any evidence of leakage shall
be cause for rejection. Following the test interval, remove the assembly from
the vessel and operate and inspect the equipment obtaining results in
accordance with 1.2.3.4.
3.12 Chemicals.
Testing is performed to determine the resistance of equipment to the vapors of
and in direct contact with various chemicals that will be encountered during
vehicle storage and service conditions.
3.12.1 General effects.
Chemicals can cause deterioration of protective coatings, damage seals, and
physically destroy parts. Damage to seals could cause leaks that could be
detrimental to personnel and hardware operation under adverse environmental
conditions. Damage to finishes can result in shortened component life.
3.12.2 Equipment specification data
required. Before proceeding, verify that the equipment specification
specifies the following as a minimum: required pretest data; failure criteria;
chemicals to be used; and duration of test if other than that specified herein.
3.12.3 Procedure. Mount the
assembly so that each of the surfaces that would be normally exposed to the
chemicals are accessible. Then expose the assembly to the vapors of and in
"direct contact" with each of the chemicals specified in the
equipment specification for a period of 48 hours minimum. By "direct
contact," it is not necessarily meant that the item need be totally
immersed in a fluid, but rather that it be covered as if by a spill of fluid
over its surfaces; reapply the fluid periodically during the 48 hour period
should evaporation cause a drying of the fluid during the chemical exposure.
After exposure, clean the external surfaces, as applicable, inspect the equipment
for physical damage, and obtain results in accordance with 1.2.3.4.
3.13 Bench
handling shock. Bench handling shock testing is performed to determine
the ability of equipment to withstand shocks that may inadvertently occur
during such handling.
3.13.1 Equipment specification data
required. Before proceeding, verify that the equipment specification
specifies pretest data required.
3.13.2 Procedures. If the
equipment is an enclosure, covers and/or panels shall be removed as if for
servicing the equipment. The assembly to be tested shall be placed in a
suitable position for servicing on a horizontal, solid wooden bench top at
least 1-5/8 inches thick. The test shall then be performed as follows:
a. Without
the equipment operating during test and using one edge as a pivot, lift the
opposite edge of the chassis until one of the following conditions occurs
(whichever occurs first):
(1) The
chassis forms an angle of 45 degrees with the horizontal bench top.
(2) The
lifted edge of the chassis has been raised 4 inches above the horizontal bench
top.
(3) The
lifted edge of the chassis is just below the point of perfect balance.
b. With
the edge lifted, let the assembly drop freely to the horizontal bench top.
Repeat the procedure using other practical edges of the same horizontal face as
pivot points for a total of four drops onto the surface.
c. Repeat
steps a and b with the test item resting on other faces until it has been
dropped for a total of four times on each face on which the test item could be
practically placed during servicing.
d. At
the conclusion of the test, inspect the item for physical damage and obtain
results in accordance with 1.2.3.4.
3.14 Temperature
shock. The temperature shock test is conducted to determine the
effects on equipment of sudden changes in temperature of the surrounding
atmosphere.
3.14.1 General effects. Adverse
effects could occur in service due to, for example, rapid altitude changes
during shipments and airdrops.
3.14.2 Equipment specification data
required. Before proceeding, verify that the equipment specification
specifies required pre-test data and failure criteria.
3.14.3 Procedure. Using a high
temperature chamber and a low temperature chamber, proceed as follows:
a. Prepare
the test item in accordance with the General Requirements of 1.2 and raise the
internal chamber temperature to 160°F. Maintain for a period of not less than
4 hours or until the test item stabilizes.
NOTE: For
steps b and c following, and when authorized by the procuring activity, large
or heavy test items shall be transferred from one chamber to the other in the
minimum practical times.
b. At
the conclusion of the previous time period, the test item shall be transferred,
within 5 minutes, to a cold chamber with an internal chamber temperature
-60°F. The test item shall be exposed to this temperature for a period of not
less than 4 hours, or until the test item stabilizes.
c. At
the conclusion of the previous time period, the test item shall, within 5
minutes, be returned to the high temperature chamber maintained at 160°F. The
test item shall be exposed to this temperature for a period of not less than 4
hours, or until the test item stabilizes.
d. Repeat
steps b and c and then go to step e.
e. Repeat
step b and then go to step f.
f. Return
the test item to standard ambient conditions and stabilize.
g. Operate
and inspect the test item and obtain results in accordance with General
Requirements of 1.2.3.4.
3.15 Explosive
atmosphere. The explosive atmosphere test is conducted to determine
the ability of equipment to operate in the presence of an explosive atmosphere
without creating an explosion or to contain an explosion occurring inside the
equipment. Since equipments operate in ever changing potentially explosive
atmospheres, the equipments, when being laboratory tested, must operate in the
presence of the optimum fuel-air mixture which requires the least amount of
energy for ignition. The equipment igniting energy may be produced electrically,
thermally, or chemically.
3.15.1 Equipment specification data
required. Before proceeding, verify the equipment specification specifies
the following as a minimum: required pretest data; failure criteria; mechanical
and/or electrical load (3.15.3.1 b); and chamber temperature condition if lower
than 160°F.
3.15.2 Apparatus. A chamber
capable of providing and verifying the explosion-proof test conditions. MIL-C-9435 describes one type of chamber that may be used.
3.15.2.1 Fuel. Unless otherwise
specified, the fuel used shall be gasoline, grade 100/130 conforming to ASTM
D910-88 or a 1.05 stoichiometric mixture of propane (C3H8)
and air within the following limits: propane 3.85 to 4.25 percent by volume
and air 96.15 to 95.75 percent by volume.
3.15.2.2 Calculation of
fuel-air-vapor ratio. As an illustration of the procedure for calculating
the weight of 100/130 octane gasoline required to produce the desired 13-to-1
air-vapor ratio, the following sample problem is presented:
a. Required
information:
Chamber air temperature during
test: 81°F.
Fuel temperature: 75°F.
Specific gravity of fuel at
61°F: 0.704.
Test altitude: 20,000 feet (P
= 6.75 lbs./in2).
Air-vapor ratio (desired): 13-to-1.
b. Employing
the following equation, calculate the apparent air-vapor ratio:
AAV =
AV (desired) = 13 =
29.70
P 6.75
1.04
14.696 -0.04 1.04 14.696
-0.04
where:
AAV = Apparent
air-vapor ratio
AV = Desired
air-vapor ratio
P =
Pressure equivalent of altitude, lbs./in2
c. At
or above 10,000 feet altitude, with chamber air temperature above 61°F and at
AV ratio of 5 or greater, air-vapor ratio = air-fuel ratio (AF) for 100/130
octane fuel. Since the conditions of the explosion test under consideration
will always be well above these values, AV will equal AF in all cases.
d. Since
AV = AF, use Figure 10 to determine weight of air (WA) and divide by AAV to
obtain uncorrected weight of fuel required (WFU).
WFU =
WA = 3.455 = 0.116 lbs., fuel weight
29.68
29.68 (uncorrected).
Note: Figure 10
pertains to a specific test chamber and shall not be used for all test facilities.
It is utilized herein for illustration of the method of employment only. Each
test chambermust have its own chamber volume chart.
e. Knowing
fuel temperatures and specific gravity at 61°F, use Figure 11 to determine
specific gravity at given temperature.
f. Using
Figure 12, obtain correction factor K for the specific gravity determined in
step e preceding. Apply factor to obtain weight of fuel corrected (WFC).
WFC = KWFU
= 1.01 x 0.116 = 0.117 lbs, fuel weight (corrected).

g. The
equipment used to vaporize the fuel for use in the explosion-proof test should
be so designed that a small quantity of air and fuel vapor will be heated
together to a temperature such that the fuel vapor will not condense as it is
drawn from the vaporizer into the chamber.
3.15.3 Procedures
3.15.3.1 Preparation for test.
The test item shall be prepared in accordance with general requirements of 1.2
and in such a manner that normal electrical operation is possible and
mechanical controls may be operated through the pressure seals from the
exterior of the chamber. External covers of the test item shall be removed or
loosened to facilitate the penetration of the explosive mixture. Large test
items may be tested, one or more units at a time, by extending electrical
connections through the cable port to the balance of the associated equipment
located externally.
a. The
test item shall be operated to determine that it is functioning properly and to
observe the location of any sparking or high temperature components which may
constitute potential explosion hazards.
b. Mechanical
loads on drive assemblies and servomechanical and electrical loads on switches
and relays may be simulated when necessary if proper precaution is given to
duplicating the normal load in respect to torque, voltage, current, inductive
reactance, etc. In all instances, it shall be considered preferable to operate
the test item as it normally functions in the system during service use.
3.15.3.2 Performance of test.
Except as specified below, the test shall be conducted as follows at simulated
test altitudes of sea level to 5,000 feet, 10,000 feet and 15,000 feet above
sea level. If the item causes an explosion at any of the test altitudes, it
shall have been considered to have failed the test and no further trials need
be attempted.
a. The
test chamber shall be sealed and the ambient temperature within shall be raised
to 160°F ± 5°F, or to the maximum temperature to which the test item is
designed to operate if lower than 160°F. The temperature of the test item and
the chamber walls shall be permitted to rise to within 20°F of that of the
chamber ambient air prior to introduction of the explosive mixture.
b. The
internal test chamber pressure shall be reduced sufficiently to simulate an
altitude approximately 10,000 feet above the desired test altitude. The weight
of fuel necessary to produce an air-vapor ratio of 13-to-1 at the desired test
altitude shall be determined from consideration of chamber volume, fuel temperature
and specific gravity, chamber air and wall temperature, test altitude, etc. If
the performance of this test will expose the equipment to an altitude in excess
of its specified maximum operating or nonoperating altitude, the next lower
simulated test altitude shall be selected. The testing of ground equipment
(equipment for operational use on the ground or aboard ship) shall be limited
to an increase in the chamber altitude to 15,000 feet with the equipment
nonoperating and the weight of fuel shall be admitted to the chamber while the
altitude is being reduced to 10,000 feet (see 3.15.2.2). A time of 3 ± 1
minutes shall be allowed for introduction and vaporization of the fuel. Air
shall be admitted into the chamber until a simulated altitude of 5,000 feet
above the test altitude is attained.
c.
Operation of the test item shall
start at this time with all making and breaking electrical contacts being
actuated. If high temperature components are present, a warmup time of 15
minutes shall be permitted. If no explosion results, air shall be admitted
into the chamber so as to steadily reduce the altitude down past the desired
test altitude to an elevation 5,000 feet below that altitude. If necessary,
conduct tests to simulated sea level pressure by pressurizing the chamber.
Tests shall not be conducted below sea level. The operation of the test item
shall be continuous throughout this period of altitude reduction and all making
and breaking electrical contacts shall be operated as frequently as deemed practicable.
Ground equipment shall be energized at 15,000 feet and the sources of igniting
energy operated until the chamber altitude is reduced to sea level.
d. If no
explosion has occurred as a result of operation of the test item by the time
the simulated altitude has been reduced to 5,000 feet below the test altitude
(or sea level in the case of ground equipment), the potential explosiveness of
the air-vapor mixture shall be verified by igniting a sample of the mixture
with a spark gap or glow plug. At any altitude below 20,000 feet, the mixture
sample shall ignite immediately at the point within 3,000 feet of the test
altitude. If the air-vapor mixture is not found to be explosive, the test
shall be considered void and the entire procedure repeated.

